Table T1. XRF
analytical conditions for Leg 179.
Oxide
or element |
Line |
Crystal |
Detector |
Collimator |
Peak angle
(°2) |
Background
offset
(°2) |
Count time
on peak
(s) |
Count time
on
background
(s) |
Analytical
error
(rel %) |
Detection
limit |
SiO2
|
K
|
PET
|
FPC
|
Medium
|
109.21
|
|
40
|
|
0.30
|
0.03
|
TiO2
|
K
|
LIF200
|
FPC
|
Fine
|
86.14
|
|
40
|
|
0.37
|
0.01
|
Al2O3
|
K
|
PET
|
FPC
|
Medium
|
145.12
|
|
100
|
|
0.40
|
0.01
|
Fe2O3
|
K
|
LIF200
|
FPC
|
Fine
|
57.52
|
|
40
|
|
0.20
|
0.01
|
MnO
|
K
|
LIF200
|
FPC
|
Fine
|
62.97
|
|
100
|
|
0.10
|
0.005
|
MgO
|
K
|
TLAP
|
FPC
|
Medium
|
45.17
|
±0.80
|
150
|
150
|
0.60
|
0.01
|
CaO
|
K
|
LIF200
|
FPC
|
Medium
|
113.09
|
|
40
|
|
0.30
|
0.005
|
Na2O
|
K
|
TLAP
|
FPC
|
Medium
|
54.10
|
-1.20
|
150
|
150
|
3.8
|
0.03
|
K2O
|
K
|
LIF200
|
FPC
|
Medium
|
136.69
|
|
100
|
|
0.40
|
0.01
|
P2O5
|
K
|
GE111
|
FPC
|
Medium
|
141.04
|
|
100
|
|
0.40
|
0.01
|
Rh
|
K Compton
|
LIF200
|
Scint
|
Fine
|
18.58
|
|
60
|
|
|
|
Nb
|
K
|
LIF200
|
Scint
|
Fine
|
21.40
|
+0.35
|
200
|
100
|
0.2
|
0.5
|
Zr
|
K
|
LIF200
|
Scint
|
Fine
|
22.55
|
-0.35
|
100
|
50
|
0.5
|
0.6
|
Y
|
K
|
LIF200
|
Scint
|
Fine
|
23.80
|
-0.40
|
100
|
50
|
1.0
|
0.6
|
Sr
|
K
|
LIF200
|
Scint
|
Fine
|
25.15
|
-0.40
|
100
|
50
|
0.3
|
0.6
|
Rb
|
K
|
LIF200
|
Scint
|
Fine
|
26.62
|
-0.60
|
100
|
50
|
3.6
|
0.6
|
Zn
|
K
|
LIF200
|
Scint
|
Medium
|
41.81
|
-0.55
|
100
|
50
|
1.0
|
1.0
|
Cu
|
K
|
LIF200
|
Scint
|
Fine
|
45.03
|
-0.55
|
100
|
50
|
1.5
|
1.2
|
Ni
|
K
|
LIF200
|
Scint
|
Medium
|
48.67
|
-0.60
|
100
|
50
|
0.6
|
1.2
|
Cr
|
K
|
LIF200
|
FPC
|
Fine
|
69.35
|
-0.50
|
100
|
50
|
1.3
|
2.0
|
V
|
K
|
LIF220
|
FPC
|
Fine
|
123.06
|
-0.50
|
100
|
50
|
2.0
|
3.0
|
Ti
|
K
|
LIF200
|
FPC
|
Fine
|
86.14
|
+0.50
|
40
|
20
|
|
|
Ce
|
L
|
LIF220
|
FPC
|
Medium
|
128.13
|
-1.50
|
100
|
50
|
|
|
Ba
|
Lß
|
LIF220
|
FPC
|
Medium
|
128.78
|
-1.50
|
100
|
50
|
|
|
Notes: All major elements measured
using a Rhodium X-ray tube operated at 30 kV and 80 mA. Trace elements are measured using
a Rhodium X-ray tube operated at 60 kV and 50 mA. Detector: FPC = flow proportional
counter (P10 gas); Scint = NaI Scintillation counter. Detection limit units = weight
percent for major elements, parts per million for trace elements.