Table T1. XRF analytical conditions for Leg 179.

Oxide
or element

 

Line

 

Crystal

Detector

Collimator

 

Peak angle
(°2)

 

Background
offset
(°2)

 

Count time
on peak
(s)

 

Count time on
background
(s)

Analytical
error
(rel %)

Detection
limit

SiO2

K

PET
FPC
Medium
109.21

40

0.30
0.03
TiO2

K

LIF200
FPC
Fine
86.14

40

0.37
0.01
Al2O3

K

PET
FPC
Medium
145.12

100

0.40
0.01
Fe2O3

K

LIF200
FPC
Fine
57.52

40

0.20
0.01
MnO

K

LIF200
FPC
Fine
62.97

100

0.10
0.005
MgO

K

TLAP
FPC
Medium
45.17
±0.80
150
150
0.60
0.01
CaO

K

LIF200
FPC
Medium
113.09

40

0.30
0.005
Na2O

K

TLAP
FPC
Medium
54.10
-1.20
150
150
3.8
0.03
K2O

K

LIF200
FPC
Medium
136.69

100

0.40
0.01
P2O5

K

GE111
FPC
Medium
141.04

100

0.40
0.01
Rh

K Compton

LIF200
Scint
Fine
18.58

60



Nb

K

LIF200
Scint
Fine
21.40
+0.35
200
100
0.2
0.5
Zr

K

LIF200
Scint
Fine
22.55
-0.35
100
50
0.5
0.6
Y

K

LIF200
Scint
Fine
23.80
-0.40
100
50
1.0
0.6
Sr

K

LIF200
Scint
Fine
25.15
-0.40
100
50
0.3
0.6
Rb

K

LIF200
Scint
Fine
26.62
-0.60
100
50
3.6
0.6
Zn

K

LIF200
Scint
Medium
41.81
-0.55
100
50
1.0
1.0
Cu

K

LIF200
Scint
Fine
45.03
-0.55
100
50
1.5
1.2
Ni

K

LIF200
Scint
Medium
48.67
-0.60
100
50
0.6
1.2
Cr

K

LIF200
FPC
Fine
69.35
-0.50
100
50
1.3
2.0
V

K

LIF220
FPC
Fine
123.06
-0.50
100
50
2.0
3.0
Ti

K

LIF200
FPC
Fine
86.14
+0.50
40
20


Ce

L

LIF220
FPC
Medium
128.13
-1.50
100
50


Ba


LIF220
FPC
Medium
128.78
-1.50
100
50


Notes: All major elements measured using a Rhodium X-ray tube operated at 30 kV and 80 mA. Trace elements are measured using a Rhodium X-ray tube operated at 60 kV and 50 mA. Detector: FPC = flow proportional counter (P10 gas); Scint = NaI Scintillation counter. Detection limit units = weight percent for major elements, parts per million for trace elements.