Table T3. XRF analytical conditions for Leg 180.

Oxide or element Line Crystal Detector Collimator Peak
angle
(° 2)
Background
offset
(° 2)
Count time (s) Analytical
error
(relative %)
Detection
limit
Peak Background

SiO2

K

PET

FPC

Medium

109.21


40


0.30

0.03

TiO2

K

LIF200

FPC

Fine

86.14


40


0.37

0.01

Al2O3

K

PET

FPC

Medium

145.12


100


0.40

0.01

Fe2O3

K

LIF200

FPC

Fine

57.52


40


0.20

0.01

MnO

K

LIF200

FPC

Fine

62.97


100


0.10

0.005

MgO

K

TLAP

FPC

Medium

45.17

±0.80

150

150

0.60

0.01

CaO

K

LIF200

FPC

Medium

113.09


40


0.30

0.005

Na2O

K

TLAP

FPC

Medium

54.10

-1.20

150

150

3.8

0.03

K2O

K

LIF200

FPC

Medium

136.69


100


0.40

0.01

P2O5

K

GE111

FPC

Medium

141.04


100


0.40

0.01























Rh

K
Compton

LIF200

Scint

Fine

18.58


60




Nb

K

LIF200

Scint

Fine

21.40

+0.35

200

100

0.2

0.5

Zr

K

LIF200

Scint

Fine

22.55

-0.35

100

50

0.5

0.6

Y

K

LIF200

Scint

Fine

23.80

-0.40

100

50

1.0

0.6

Sr

K

LIF200

Scint

Fine

25.15

-0.40

100

50

0.3

0.6

Rb

K

LIF200

Scint

Fine

26.62

-0.60

100

50

3.6

0.6

Zn

K

LIF200

Scint

Medium

41.81

-0.55

100

50

1.0

1.0

Cu

K

LIF200

Scint

Fine

45.03

-0.55

100

50

1.5

1.2

Ni

K

LIF200

Scint

Medium

48.67

-0.60

100

50

0.6

1.2

Cr

K

LIF200

FPC

Fine

69.35

-0.50

100

50

1.3

2.0

V

K

LIF220

FPC

Fine

123.06

-0.50

100

50

2.0

3.0

Ti

K

LIF200

FPC

Fine

86.14

+0.50

40

20



Ce

L

LIF220

FPC

Medium

128.13

-1.50

100

50



Ba


LIF220

FPC

Medium

128.78

-1.50

100

50


Notes: All major elements were measured using a rhodium X-ray tube operated at 30 kV and 80 mA. Trace elements were measured using a rhodium X-ray tube operated at 60 kV and 50 mA. Detector: FPC = flow proportional counter (P10 gas); Scint = NaI scintillation counter.