Table T5. Instrumental conditions for XRF major and trace element standard measurements.

Element Line Crystal Detector Collimator Peak
angle
(°2)
Background offset
(°2)
Total
peak
(s)
Count
time background (s)
Analytical error
(SD%)
Detection limit
(%)
Major









SiO2 SiKa1,2r PET FPC Medium 109.21 NB 60 NB 0.20 0.030
TiO2 TiKa1,2r LIF200 FPC Fine 86.14 NB 100 NB 0.56 0.009
Al2O3 AlKa1,2r PET FPC Medium 145.12 NB 100 NB 0.55 0.041
Fe2O3 FeKa1,2r LIF200 FPC Fine 57.52 NB 90 NB 0.42 0.007
MnO MnKa1,2r LIF200 FPC Fine 62.97 NB 200 NB 2.12 0.006
MgO MgKa1,2 TLAP FPC Medium 45.17 -0.80 200 200 0.40 0.050
CaO CaKa1,2r LIF200 FPC Medium 113.09 NB 120 NB 0.25 0.003
Na2O NaKa1,2 TLAP FPC Medium 55.10 1.20 210 210 1.21 0.100
K2O KKa1,2r LIF200 FPC Medium 136.69 NB 175 NB 0.89 0.001
P2O5 PKa1,2s GE111 FPC Medium 141.04 NB 175 NB 1.56 0.003
Trace









Nb Ka LIF200 Scint Fine 21.40 +0.35 100 100 2.0 1
Zr Ka LIF200 Scint Fine 22.55 -0.35 100 100 0.7 1
Y Ka LIF200 Scint Fine 23.80 -0.40 100 100 2.7 1
Sr Ka LIF200 Scint Fine 25.15 -0.40 100 100 1.1 1
Rb Ka LIF200 Scint Fine 26.62 -0.60 100 100 8.1 1
Zn Ka LIF200 Scint Fine 41.81 -0.55 100 100 1.8 2
Cu Ka LIF200 Scint Fine 45.03 -0.40 100 100 1.3 2
Ni Ka LIF200 Scint Fine 48.67 -0.40 100 100 1.2 2
Cr Ka LIF200 FPC Fine 69.35 -0.50 100 100 0.8 3
V Ka LIF220 FPC Fine 123.06 -0.50 100 100 1.1 4
Ce La LIF220 FPC Medium 128.16 -1.50 100 100 5.6 10
Ba Kb LIF220 FPC Medium 128.78 -1.50 100 100 5.8 15

Notes: SD = standard deviation; NB = no background correction. The standard reference material BHVO-1 was used to quantify analytical error.