Table T6. JY2000 ICP-AES operating conditions and analytical sample run parameters, Leg 187.

Sample designation Leg
187A
Leg
187B
Leg
187C
Leg
187D
Leg
187E
Leg
187F
Leg
187G
Leg
187H
Leg
187I
Leg
187J
Leg
187L
Date (1999): 30 Nov 99 2 Dec 99 3 Dec 99 8 Dec 99 13 Dec 99 14 Dec 99 17 Dec 99 19 Dec 99 26 Dec 99 1 Jan 00 5 Jan 00
Nebulizer type: V-Groove V-Groove V-Groove V-Groove Concentric Concentric Concentric Concentric Concentric Concentric Concentric
Sample dilution factor: 505 505 505 505 2005 2005 2005 2005 2005 2005 2005
Pump speed (mL/min): 20 (1) 30 (1.5) 30 (1.5) 30 (1.5) 23 (1.2) 23 (1.2) 23 (1.2) 23 (1.2) 23 (1.2) 23 (1.2) 23 (1.2)
Uptake time (s): 40 40 40 40 40 40 40 40 40 40 40
Rinse time (s): 90 90 90 90 90 90 90 90 90 90 90
Power setting (kv): 1000 1000 1000 1000 1000 1000 1050 1050 1050 1050 1050
Ar humidifier: Yes Yes Yes Yes No No No No No No No
Nebulizer pressure (bars): 2.5 2.75 2.94 2.94 3 2.75 3 3 3 3 3
Nebulizer flow (starting-ending) (L/min): ? ? 0.64 0.64 0.83-0.41 0.82 0.84 0.83-0.38 0.75-0.60 ? ?
G-3 gas flow (L/min): 1 1 0.8 (reset) 0.8 0.8 0.8 (reset) 0.8 0.8 0.8 0.8 0.8
Method conditions










Mode: 5 5 5 5 5 5 5 5 5

Number of replicates: 3 3 3 3 3 3 3 3 3

Counting time (s): 1 1 1 1,(2-Cr,Ni,Na,K) 1,(2-Cr,Ni,Na,K) 1,(2-Cr,Ni,Na,K) 1,(2-Cr,Ni,Na,K) 1,(2-Cr,Ni,Na,K) 1,(2-Cr,Ni,Na,K) 1,(2-Cr,Ni,Na,K) 1,(2-Cr,Ni,Na,K)

Notes: ? = not recorded. This table is also available in ASCII format.