Table T1. Analytical information from Leg 199.
Element
Wavelength
(nm)
Detection
limit (wt%
or ppm)
Within-run
drift
(%)**
Within-run
precision
(%)††
Within-site
precision
(%)‡‡
Long-term
precision
(%)***
Si 251.611 0.1 1.2 1.7 0.9 3.3
Al 396.152 0.02 1.1 0.8 3.8 5.1
Ti 334.941 0.002 0.5 0.6 1.3 21.7
Fe 259.940 0.003 2.9 0.9 2.4 9.2
Mn 257.610 0.001 0.8 1.2 1.9 8.1
Ca 393.366 0.012 1.7 0.9 0.7 11.5
Mg 285.213 0.003 0.5 7.6 2.3 12.7
P* 178.229 0.07 30.8 68.5 67.8 32.9
P 178.229 0.07 11.2 8.6
Ba 455.403 3.9 1.7 0.3 2.8 21.3
Sr 407.771 1.7 1.3 1.0 1.9 14.7

Notes: * = phosphorus analyses using the nitrogen generator, Site 1218. † = phosphorous analyses using the nitrogen storage tank, Site 1221. ‡ = detection limits calculated as three times the standard deviation of the blank: Si, Al, Ti, Fe, Mn, Ca, Mg, and P listed in weight percent, and Ba and Sr listed in ppm. ** = representative drift values within a single analytical run, Site 1218. †† = representative precision for a single typical analytical run using JCh-1, Site 1218. ‡‡ = representative precision using drift solution, Site 1218. *** = based on precision values, Sites 1215-1221.