Table T10. Interstitial water analyses and corresponding reproducibility.
Parameter
Method
IAPSO value
Analytical uncertainty
Accepted
Measured
1 (%)
n
Alkalinity Gran titration 2.33 2.34 2.2 10
Cl (mM) Titration with AgCl 559 544 0.7 12
H4 SiO4 (然) Spectrophotometry

0.8 5
NH4 (然) Spectrophotometry



PO4 (然) Spectrophotometry

BDL
SO4 (mM) Ion chromatography 28.9 28.9 0.3 6
Mg (mM) ICP-AES 10.55 10.63 2.1 6
Ca (mM) ICP-AES 54.0 51.64 2.2 6
K (mM) ICP-AES 10.44 10.26 4.6 6
Na (mM) ICP-AES
465 0.4 6
Sr (然) ICP-AES 87 90 5.8 6
Ba (然) ICP-AES
70 0.6 6
B (然) ICP-AES 416 546 2.9 6
Li (然) ICP-AES 27 21 5.4 6
Mn (然) ICP-AES 0 0.6 2.6 6
Fe (然) ICP-AES 0 BDL BDL
DOC (mM) Gas chromatography

9.6 3

Notes: IAPSO = International Association for the Physical Sciences of the Ocean, standard seawater. BDL = below detection limit. DOC = dissolved organic carbon. ICP-AES = inductively coupled plasma-atomic emission spectroscopy. n = number of replicate analyses. This table is also available in ASCII.