Parameter
|
Setting
|
SQUID configuration: |
|
SQUID-x response (cm) | 6.071 |
SQUID-y response (cm) | 6.208 |
SQUID-z response (cm) | 9.923 |
SQUID-x calibration (emu) | 8.21E-05 |
SQUID-y calibration (emu) | -8.34E-05 |
SQUID-z calibration (emu) | 4.32E-05 |
SQUID filter data (x, y, and z) (Hz) | 10 |
SQUID range data (x, y, and z) | 1 x range |
SQUID slew data (x, y, and z) | Disable fast slew |
SQUID feedback data (x, y, and z) | Close feedback loop |
Degausser configuration: |
|
Ramp rate | 3 |
Tracking timeout (s) | 40 |
AF field maximum (mT) | 80 |
AF field minimum (mT) | 2 |
Coil calibration constant (x, y, and z) | 1 |
Character delay | 50 |
Sample handler configuration (right-hand home and right-hand tray): |
|
Home switch type | Limit switch |
Home offset (cm) | 0.0 |
Tray offset (cm) | 2.2 |
Track length (cm) | 599.3 |
Full tray length (cm) | 160.3 |
Scale (steps/cm) | 1572 |
Sample handler positions: |
|
SQUID (cm) | 255.85 |
Magnetic susceptibility (not applicable) | 0 |
Degauss stage 1 (cm) | 190.0 |
Degauss stage 2 (cm) | 435.0 |
DG coil center (cm) | 410.0 |
IRM stage (not applicable) | 0 |
IRM coil center | 0 |
Background stage 1 (cm) | 300.0 |
Background stage 2 (cm) | 30.0 |
Load/unload (cm) | 0.0 |
Sensor velocity (cm/s) | 10-25 |
Degauss velocity (cm/s) | 10-25 |
Background velocity (cm/s) | 10-25 |
Load velocity (cm/s) | 10-25 |
Discrete configuration: |
|
Max number of sample positions | 4 |
Center offset (cm) | 20 |
Center separation (cm) | 40 |
Notes: SQUID = super-conducting quantum interference device. AF = alternating field. IRM = isothermal remanent magnetization.