Table T10. Reproducibility for analytical determinations, Leg 207.
Parameter
Method
Reproducibility
Alkalinity Gran titration ±1.5%
Cl Titration with AgNO3 ±0.1%
K+ ICP-AES ±0.02%
Mg2+ ICP-AES ±0.1%
Ca2+ ICP-AES ±0.2%
SO42– Ion chromatography ±0.1%
Si(OH)4 Spectrophotometry ±2 mM
NH4+ Spectrophotometry ±4 mM
HPO42– Spectrophotometry ±0.2 mM
Fe2+ ICP-AES ±2.3%
Mn2+ ICP-AES ±2.3%
B ICP-AES ±2.4%
Li+ ICP-AES ±4.4%
Sr2+ ICP-AES ±1.0%
Ba2+ ICP-AES ±1.1%

Note: ICP-AES = inductively coupled plasma–atomic emission spectroscopy.