Initial Reports, Volume 209 Ms 209IR-102 Chapter 2, Table T8. Estimates of accuracy and precision for Carlo Erba gas chromatography analyses of CO2, H2O, and S. Notes: Preferred values: BAS140 (Sparks and Zuleger, 1995), JP-1 (Sims et al., 1988), and 147-895D-10W (Puchelt et al, 1996). NA = not applicable (no preferred value). Standard Date (2003) Volatiles: CO2 (wt%) H2O (wt%) S (ppm) JP-1 23 May 0.31 3.04 27 May 2.39 28 May 0.30 2.94 31 May 0.29 2.73 5 Jun 0.30 2.95 10 Jun 0.33 3.05 13 Jun 0.33 3.21 16 Jun 0.32 2.43 21 Jun 0.36 2.32 Average (wt%): 0.32 2.79 Preferred value (wt%): 3 Standard deviation (wt%): 0.02 0.33 Precision (%): 7.00 11.74 Accuracy (%): NA 0.55 BAS140 23 May 0.12 1.20 841 27 May 1.42 28 May 0.12 1.05 1237 31 May 0.11 1.06 929 5 Jun 0.11 1.12 546 10 Jun 0.13 1.16 1479 13 Jun 0.11 1.26 628 16 Jun 0.13 1.93 493 21 Jun 0.16 1.78 581 Average (wt%): 0.12 1.33 842 Preferred value (wt%): 0.05 1.12 900 Standard deviation (wt%): 0.02 0.32 357 Precision (%): 12.86 24.00 42.42 Accuracy (%): 58.79 15.90 -6.93 147-895D-10W 23 May 0.98 10.76 27 May 0.88 10.43 28 May 1.08 11.03 31 May 1.05 11.33 5 Jun 0.92 11.66 10 Jun 1.41 11.36 13 Jun 1.09 13.78 16 Jun 0.97 9.99 21 Jun 1.14 11.14 Average (wt%): 1.06 11.28 Preferred value (wt%): 0.95 9.62 600 Standard deviation (wt%): 0.16 1.07 Precision (%): 14.84 9.48 Accuracy (%): 11.24 17.23