BASIC OVERVIEW OF ICP-AES

ICP-AES is an emission spectrophotometric technique, exploiting the fact that excited electrons emit energy at a given wavelength as they return to ground state. The fundamental characteristic of this process is that each element emits energy at specific wavelengths peculiar to its chemical character. Although each element emits energy at multiple wavelengths, in the ICP-AES technique it is most common to select a single wavelength (or a very few) for a given element. The intensity of the energy emitted at the chosen wavelength is proportional to the amount (concentration) of that element in the analyzed sample. Thus, by determining which wavelengths are emitted by a sample and by determining their intensities, the analyst can quantify the elemental composition of the given sample relative to a reference standard.

ICP-AES analysis requires a sample to be in solution. Thus, interstitial waters can be analyzed simply, requiring only dilution in most cases. Igneous rocks, sedimentary rocks, and sediments, however, must be dissolved. This can be achieved either by a combined acid attack employing HF, HNO3, and HCl acids, or by a LiBO2 flux-fusion technique similar to that used for XRF preparation. In addition to being somewhat dangerous for routine shipboard work (because of the highly reactive nature of HF), the acid attack is not able to generate consistent and reliable data for Si because it volatilizes in the presence of HF. The acid digestion procedure also often results in incomplete analysis of refractory elements such as Ti, Cr, and Zr because their host minerals are often difficult to dissolve. The flux-fusion approach is employed on board the Resolution for several reasons: (1) it is safer because HF is not involved; (2) it is a complete dissolution technique, allowing determination of all elements, including Si and the refractory elements; (3) the resultant solutions are similar in composition (or matrix) because they are dominated by the presence of the LiBO2 flux; and (4) the solutions are stable in dilute HNO3 acid and can be transported safely back to the scientist's laboratory for further shore-based study.

All ICP-AES systems consist of several components. We shall focus this discussion on three main aspects: the sample introduction system, the torch assembly, and the spectrometer. The sample introduction system on the ICP-AES consists of a peristaltic pump, Teflon tubing, a nebulizer, and a spray chamber. The fluid sample is pumped into the nebulizer via the peristaltic pump. The nebulizer generates an aerosol mist and injects humidified Ar gas into the chamber along with the sample. See "Nebulizers," for a more detailed discussion of nebulizers. This mist accumulates in the spray chamber, where the largest mist particles settle out as waste and the finest particles are subsequently swept into the torch assembly. Approximately 1% of the total solution eventually enters the torch as a mist, whereas the remainder is pumped away as waste.

Humidification of the Ar gas injected into the nebulizer is important when analyzing samples with high dissolved solids, as is often the case with analysis of ODP rocks, sediments, and interstitial waters. Humidification takes place in the Ar humidifier, where Ar is bubbled through deionized water prior to its expulsion in the nebulizer.

The fine aerosol mist containing Ar gas and sample is injected vertically up the length of the torch assembly into the plasma. There are several recommended Ar flow rates used in the torch, as described in detail in the owner's manual and in the various publications provided. The radio frequency-generated and maintained Ar plasma, portions of which are as hot as 10,000 K, excites the electrons. When the electrons return to ground state at a certain spatial position in the plasma, they emit energy at the specific wavelengths peculiar to the sample's elemental composition.

The plasma is viewed horizontally by an optical channel. Light emitted from the plasma is focused through a lens and passed through an entrance slit into the spectrometer. There are two types of spectrometers used in ICP-AES analysis: sequential (monochromator) and simultaneous (polychromator). The JY2000 has a sequential spectrometer. This means that the diffraction grating in the spectrometer is analogous to a prism that refracts visible light into its component colors. The detector (photomultiplier tube) is fixed in space at the far end of the spectrometer. Rotation of the diffraction grating sequentially moves each wavelength into the detector. The computer control ensures that the detector is synchronized with the grating so that the intensity at the detector at any given time is correlated with the wavelength being diffracted by the grating. The operator enters the wavelengths that he or she wishes to detect into the computer, the grating sequentially moves to the specified wavelengths, and the energy intensity at each wavelength is measured to provide a quantitative result that can be compared to a reference standard. Using standard spectroscopic techniques (e.g., background corrections), sequential ICP-AES can provide extremely flexible and rapid analysis of a number of chemical elements. The spectrometer is flushed with N2 gas to improve the detection limits of elements with emission wavelengths that are severely compromised by interference with air (e.g., P). This N2 flush, which is constantly maintained in the instrument regardless of whether such elements are being analyzed, also protects the optics from the corrosive aspects of the atmosphere, which are particularly acute at sea.