MATERIALS AND METHODS

Chemical Tracer: Perfluorocarbon

Characteristics

Perfluorocarbon tracers (PFT) have been used extensively in land-based drilling applications (Senum and Dietz, 1991; Russell et al. 1992; McKinley and Colwell, 1996) because they are inert and can be detected with high sensitivity. Perfluoro(methylcyclohexane) is the tracer that has been tested on the JOIDES Resolution. This perfluorocarbon (Aldrich 30293-7) has a molecular weight of 350.05, a boiling point of 76°C, and a density of 1.76 g/mL. Its solubility is ~1 mg/L in water and is 10 g/L in methanol (Colwell et al., 1992). The low solubility in water facilitates gas phase partitioning and quantitative headspace analysis.

Preparation and Delivery

The stock PFT is shipped in sealed ampoules, and it is not necessary to dilute it prior to use. Because the PFT is volatile and can be detected at extremely low concentrations, it is necessary to open the ampoules and transfer the PFT to the carboy used for delivery in a ventilated area well away from the core handling and PFT detection areas. Gloves should be worn during this process and discarded afterward. It is recommended that this transfer be performed on the helicopter deck while the JOIDES Resolution is under way and that all materials that may have been in contact with the PFT be disposed of immediately. These precautions will minimize the probability of obtaining false positives when analyzing core material.

The PFT is introduced into the stream of drilling fluid at a rate calculated to achieve a nominal final concentration of 1 mg/L. The injection rate is controlled with a single-piston high performance liquid chromatography (HPLC) pump (Alltech model 301), which pumps the PFT from a polypropylene carboy into the drilling mud stream through a valve on the low-pressure side of mud charge pump No. 2 (Fig. F1). It is important that the carboy is secured. An additional HPLC pump connected to mud charge pump No. 1 allows the delivery of the tracer from both mud pumps. The time needed for the tracer to reach the bit is dependent on both the pipe length and the pumping rate of the drilling fluid and can be determined using the data in Table T1. Pumping of the tracer must be started early enough to allow it to reach the bit before coring begins. The pumping rate of the mud pump typically varies from 20 to 95 strokes/min (392-1862 L/min) and is controlled by the driller. PFT injection rates are adjusted to the mud pump output to maintain the 1 mg/L concentration in the entire drill string (Table T1). Currently, the PFT injection rate is controlled manually. Plans to control the PFT injection rate from the drillers' shack are under way.

Collection of Samples

Sediments

Unconsolidated sediment is sampled on the catwalk immediately after cores are retrieved. After the core liner is cut, the sediment core should be broken by pulling the sections apart instead of using the usual procedure of cutting the sections apart with a knife. This will ensure that the tracer is not dragged through the core with the knife. Two plugs (~3 cm3 each) from the bottom end of each section are taken with 5-mL syringes that have been modified by cutting off the luer-lock end. New syringes are used for each sample. One sample is taken from the outer edge along the core liner and the other from near the center of the core. The samples are immediately extruded into 20-mL-capacity headspace vials (Hewlett Packard #5182-0837) and sealed with gas-tight caps containing septa. The sample from the outer edge is used to confirm the successful delivery of the tracer, whereas the interior sample is used to quantify the intrusion of drill water into the core.

If the sediment is too hard to sample with a syringe, the exterior of the core must first be pared away using a hammer, chisel, and tongs. To prevent contamination of the interior while paring, the tools should be passed through the flame of a torch to remove any PFT. This procedure should be done on a new sheet of aluminum foil to prevent cross contamination between samples. Samples from the exterior and interior of the core are placed in headspace vials and sealed. Because the exterior of the core liner is coated with drilling fluid, contact with the liner is to be avoided while collecting core samples for PFT analysis.

Igneous Rock

Immediately after the core liner is split in half in the core lab, pieces of the core are chosen for PFT analysis. To confirm that the PFT reached the core, several small pieces of rock are placed directly into the headspace vials (ID = 19 mm; height = 75 mm) and immediately sealed. Alternatively, a cotton swab can be used to wipe the interior of the core liner and then placed in a headspace vial and sealed.

Ideally, the PFT is removed from the surface of the rock prior to sampling the interior. This will prevent the transfer of the tracer into the interior during sample preparation. This can be accomplished by rinsing the exterior with water or methanol and then drying it under a flame. Alternatively, the rock can be dried directly under a flame without rinsing. The piece is held with tongs under the flame from a handheld propane torch until it appears dry. Experiments conducted during Leg 185 using all three methods showed that drying the surface of the rock with a flame was the best method. There is a concern that this may compromise the usefulness of the sample for subsequent microbiological samples because of the heat and/or desiccation.

After removing the PFT from the exterior, pieces from the interior of the rock are obtained by paring away the exterior using a hammer and chisel while the rock is held on a sheet of aluminum foil. This may be accomplished more effectively using the hydraulic rock splitter. Tongs are used to move the rock onto new sheets of aluminum foil. After each paring, the tools are cleaned of PFTs by passing the tools through the flame of the torch, and the rock pieces are placed on new foil sheets. When the entire exterior of the rock is removed, the residual rock (interior) is placed in a percussion mortar and crushed. Aliquots of the crushed rock are then placed in headspace vials and sealed.

Perfluorocarbon Analysis

A Hewlett Packard 8059 gas chromatograph (GC) with an electron capture detector (HP G1223A) is used to quantify the PFT in the samples. The GC is equipped with an HP-PLOT Al2O3 "M" deactivation column (length = 15 m; ID = 0.53 mm; coating thickness = 15 µm), and nitrogen is used as the carrier gas. Hewlett Packard ChemStation software (Rev. A.05.02 [273]) is used to control the GC, including acquisition of chromatographs and peak area integration. The parameters of the runs were modified from Colwell et al. (1992). The injector temperature is 180°C, and the initial oven temperature is 120°C. The oven temperature is ramped upward immediately after injection at a rate of 20°C/min to a final temperature of 150°C and held for 3 min. This results in a total run time of 4.5 min with the PFT peak eluting at ~2.7 min. The injector pressure is set at 3.0 psi, and the flow rate is 9.4 mL/min. The gas stream is not split before it reaches the detector. Samples, standards, and blanks are injected manually using Hamilton gas-tight syringes (50, 500, and 5,000 µL). To minimize adsorption of the PFT on the interior surface of the syringe, the syringe is heated to 70°C prior to injection.

Calibration

Perfluoro(methylcyclohexane) is diluted in methanol to 10-1, 10-3, 10-5, 10-7, and 10-9 vol:vol in headspace vials to use as standards. These vials are immediately sealed to minimize loss of the PFT by volatilization. A 10-µL aliquot of each solution in the dilution series is added to separate headspace vials and sealed. These vials are heated at 70°C in an oven to volatilize the PFT. Either 5 or 0.5 mL of the headspace gas is withdrawn with a heated (70°C) gas-tight syringe and injected into the GC. The resultant signal is expanded in the vertical scale, and the perfluorocarbon peak is chosen manually. A standard curve is constructed using the peak area vs. PFT injected. Measured peak areas are linear with PFT headspace abundance over four orders of magnitude (Fig. F2). Based on duplicate analysis of standards, the relative measurement error is 20% () and the detection limit is ~6 × 10-12 g PFT. This is equivalent to 0.006 µL of drilling fluid when the PFT concentration in the drilling fluid is 1 mg/L. Further details on calibrating the GC are presented in "Appendix A".

Sediment and Igneous Rock

The headspace vials containing the samples, (sediment, whole rock, or crushed rock) and the gas-tight syringes are heated to 70°C in an oven. Five mL of headspace gas is injected into the GC. The signal is expanded in the vertical scale, and the perfluorocarbon peak is chosen manually based on the elution time of the standard. The peak area is integrated and converted to the amount of PFT using the values from the standard curve. The amount of sample is determined by weighing each vial and subtracting the weight of an empty vial. The total headspace volume is calculated by subtracting the volume of sample from the total volume of the vial. The total tracer concentration in the sample is corrected to account for the fraction of the headspace that is injected. The amount of drilling fluid present in the sample is calculated assuming that the tracer was present at 1 mg/L (or the final concentration of PFT if it is known from the injection and pumping rates to deviate from 1 mg/L) (see "Appendix B").

Blanks

Procedural blanks are analyzed to determine the instrumental and procedural backgrounds. These blanks consist of 5-mL injections of air collected in the gas-tight syringe from outside the laboratory or headspace gas from empty vials prepared at the same time and location the samples are taken.

Particulate Tracer: Fluorescent Microspheres

Characteristics

Fluorescent microspheres of similar size to the indigenous microorganisms (0.5-1.0 µm) have successfully been used in drilling operations (Harvey et al., 1989). Yellow-green fluorescent (458 nm excitation, 540 nm emission) microspheres 0.518 (±0.01) µm in diameter (Fluoresbrite Carboxylate Microspheres; Polysciences Inc. #15700) are used as a particulate tracer. These microspheres appear bright green when observed by epifluorescence microscopy using a blue filter set (Zeiss filter set 09 or 10) (Fig. F3).

Preparation and Delivery

The desired concentration of microspheres at the point of drilling is ~1010 spheres/mL. To achieve this concentration, 2 mL of the microsphere suspension (3.78 × 1011 spheres/mL) is diluted to 40 mL with distilled water, and this solution is placed in an ultrasonic bath to disrupt aggregates. The microsphere suspension is placed in a plastic bag (Whirl-Pak; Nasco Inc.) and heat sealed.

A spacer with a channel to accommodate the top of the Whirl-Pak bag was machined to fit inside a core-catcher sleeve (Fig. F4A). The wire bail of the bag is taped into the groove in the brass spacer (Fig. F4B). The bag is draped over the core-catcher body and taped along the edges (Fig. F4C). The entire assembly is shown in Figure F4D. During APC drilling, the core barrel is lowered on a wireline, whereas the core barrel is allowed to free fall in the drill string from the drill floor to the bottom of the hole during RCB coring.

Collection of Samples

Sediments

As with the PFT samples (above), after the core liner is cut on the catwalk, the sediment core should be broken apart rather than cut so that the microspheres are not dragged through the core with the knife. While the core is still on the catwalk, toothpicks are used to collect small sediment samples from the bottom of core sections. If the cores are cut instead of broken, samples are retrieved from the interior of the core by boring into the core with successive toothpicks, using each only once. Sediment from the toothpick is removed from the toothpick by washing in 0.5 mL of distilled water. This can be done in either microcentrifuge tubes or in the wells of microplates. From each section that will be used for microbiological analysis, a sample is taken adjacent to the interior of the core liner, from the center of the core, and/or midway between the interior and the exterior of the core.

Igneous Rock

After splitting the core liner in the core lab, rock pieces are selected to determine the presence or absence of fluorescent microspheres. The surface of the rock is washed with distilled water, and the rinse is collected and examined with epifluorescence microscopy to verify the successful delivery of the microspheres. Samples that will be used for microbiological assays must first have the exterior pared away as with the perfluorocarbon tracer (above). After paring the exterior, the interior of the rock is crushed in a percussion mortar. Distilled water is added to subsamples of this material in preparation for examination with the epifluorescence microscope (below).

Quantification of Microspheres

Fluorescent microspheres are quantified using a Zeiss Axiophot epifluorescence microscope outfitted with a mercury lamp (HBO 100W), a blue filter set, and a 100× Plan-NEOFLUAR oil-immersion objective. Nonfluorescent immersion oil is used for all observations (see "Appendix C").

Sediments and Igneous Rock

Aliquots (50 µL) of the sediment slurry or the crushed rock are filtered onto black, 13-mm-diameter polycarbonate filters (0.2-µm pore size) in a filtration tower. The filters are then mounted on microscope slides with a drop of nonfluorescent immersion oil and covered with a cover slip. The microspheres on the filter are then counted using the epifluorescence microscope. Microsphere abundance on the filters is determined by averaging the total number seen in at least 20 randomly selected fields of view and normalizing to the area of the field of view, area of the filter used, and the volume of the sediment slurry that was filtered. The usable area of the filter is measured with calipers, and the area of the field of view is calibrated with a stage micrometer. It is critical that the filter tower used for the filter preparation is carefully cleaned of microspheres between samples to avoid cross contamination. If the amount of sample material is not weighed, the results are reported as spheres per milliliter of sediment suspension. If the material is weighed prior to preparing the filter, the results are reported as microspheres per gram of sample. The sediment sample adjacent to the core liner is used to verify the successful delivery of the microspheres in those cores, whereas filters prepared from the wash water are used for the igneous samples.

Thin sections prepared from the rocks can be examined for the presence of fluorescent microspheres by viewing with epifluorescence microscopy. A drop of nonfluorescent immersion oil is placed directly on top of the thin section, which is then examined with the epifluorescence microscope.

NEXT